Load cell calibration

The Load cell calibration box on Settings, below Motor calibration, measures the two load cell values a datasheet can only give you nominally. Neither step moves the carriage.

The Load cell calibration panel

Force is worked out from the bridge signal as (signal − Zero Balance) × Capacity ÷ Sensitivity, all in mV/V: millivolts of bridge output per volt of excitation. A load cell rated 2 mV/V puts out 2 mV per volt of excitation at its capacity, so 10 mV at 5 V. The machine measures the signal against the excitation itself, so supply drift cancels out.

Restart after writing load cell values

The machine loads the load cell values when it starts. Both steps work from the values it started with, so after Write to device, restart the machine before reading again.

Zero and noise

Take any sample out of the grips, let the load cell hang free, and press Read zero. It reads for 3 s and reports:

A zero and noise reading

  • Zero balance: the bridge signal with nothing on the load cell, next to the saved value and the force the saved one was leaving in the readings. Use this zero puts it into Load Cell Zero Balance (mV/V) in Machine configuration; then Write to device and restart.
  • Noise: one standard deviation of the readings, in force, in nV/V and as a share of capacity, and the spread from lowest to highest. This is the smallest force the machine can tell apart.
  • still settling: the reading moved more than its noise explains while it was read. Wait (a cold load cell can take minutes) and read again. The noise figure includes the drift while it does.

Sensitivity from known weights

The load cell must hang vertically, with the weight pulling it the way a sample would.

  1. Press 1. Read unloaded.
  2. Enter the mass in Mass (kg), counting the hanger, hang it, and press 2. Read with weight.
  3. Repeat step 2 with more weights, the heaviest you have, ideally up to at least 20% of capacity. Press Read unloaded again whenever you like: each weight is measured from the most recent unloaded reading, and the panel shows how far that has moved from the first one.

Each weight is the change from the unloaded reading, so the zero balance does not have to be right first. The new sensitivity is the saved one × the force read ÷ (mass × gravity); with several weights it comes from a least-squares line through zero. Gravity (m/s²) defaults to standard gravity, 9.80665; local gravity differs by up to about 0.3%.

Sensitivity fitted from three known weights

Each weight's row shows its known force, the force read, the error against the saved sensitivity, and how far it sits off the line. Linearity is the furthest any weight sits from the line, as a share of the heaviest weight. Use this sensitivity puts the value into Load Cell Sensitivity (mV/V); then Write to device and restart. Remove drops a weight.

Flags

Flag What it means What to do
polarity The weights read as pushing If they hang the way a sample pulls, the load cell is wired the other way round; the new sensitivity flips its sign to match
light The heaviest weight is under 20% of capacity, so the line is stretched a long way to full scale Add a heavier weight
swinging A weight was still moving while it was read Remove it, let it settle, read it again
moved the reading by under 50 mN That weight barely registered Check it hangs from the load cell, and that Capacity and Sensitivity are entered