Theory of operation¶
What the machine is doing, and why the app behaves the way it does. This page is for operators: it explains the model behind the badges, banners and buttons. The pages after it go into the firmware, protocol and app internals.
What the machine is¶
MaD pulls a sample apart along one axis while measuring two things: how far the jaws have moved and how hard the sample is pulling back.
flowchart LR
subgraph computer["Your computer"]
app["MaD Control<br/>(browser)"]
end
subgraph machine["The machine"]
ctrl["Controller<br/>(Propeller 2)"]
drive["Motor drive"]
motor["Motor + screw"]
enc["Encoder"]
lc["Load cell + amplifier"]
sd["SD card"]
safety["Emergency stop chain,<br/>end stops, door"]
end
app <-->|"control port (USB serial)"| ctrl
ctrl -->|step / direction / enable| drive --> motor
motor --- enc -->|position| ctrl
lc -->|force| ctrl
ctrl <--> sd
safety --> ctrl
| Part | Job |
|---|---|
| Controller | A Propeller 2 microcontroller on the Edge board. It runs the firmware: motion, measurement, safety checks, test execution and the link to the app. |
| Motor drive and motor | The controller sends step and direction pulses; the drive turns the motor, and the screw moves the carriage. |
| Encoder | Reports where the carriage actually is. The controller treats the encoder, not the pulses it sent, as the truth. |
| Load cell | A strain-gauge bridge read by a 24-bit converter on the force gauge board. |
| SD card | Holds the test program while it runs, the recorded data, and the machine configuration. |
| Safety inputs | The emergency stop chain, an end stop at each end of travel, and the door interlock. |
| MaD Control | The browser app. It configures, commands and displays. It is not in the control loop. |
How position is controlled¶
A move, whether a jog or a step of a test, becomes a trajectory: a position that changes over time, accelerating at Acceleration Max up to the move's speed and back down to a stop.
A thousand times a second the controller:
- works out where the trajectory says the carriage should be now;
- reads the encoder to see where it is;
- sends the drive step pulses for the trajectory's speed plus a correction proportional to the error (Servo Kp, with Servo Ki for a steady error).
Two configuration values convert between the controller's world and millimetres: Servo (step/mm) for pulses out and Encoder (step/mm) for counts in. If their ratio is wrong, the correction fights the motion, which is why the drive and encoder check exists.
The gap between where the trajectory is and where the encoder is, the
following error, is also how the machine notices a jam. Too large, or too
little encoder travel for the pulses sent, and it raises
SERVO_STALL.
The Control screen shows these quantities: Machine Position (the encoder), Machine Setpoint (the trajectory, where the gantry should be right now), Machine Target (where the current move ends), and on the chart the Expected path, a straight line at the commanded speed.
How force is measured¶
The load cell's bridge puts out a voltage proportional to both the force on it and the voltage exciting it. The converter measures the output relative to the excitation, in mV/V, so supply drift cancels out. Force is then:
force = (signal − Zero Balance) × Capacity ÷ Sensitivity
Those three are in the machine configuration. Capacity and Sensitivity come from the load cell's datasheet; the load cell calibration measures Zero Balance and refines Sensitivity against known weights.
Machine and sample coordinates¶
Every measurement exists in two frames:
| Machine | Sample | |
|---|---|---|
| Position | Jaw separation | Extension since Zero length |
| Force | Absolute force on the load cell | Force since Zero force |
| Used for | The machine's own limits | The test's limits, strain and stress |
Zeroing changes only the sample frame. That is why you zero with the sample mounted and just slack: the sample frame then starts at the gauge length with no load.
The state machine¶
The controller is always in one of four states, and all motion is gated on them.
stateDiagram-v2
[*] --> DISABLED
DISABLED --> MANUAL: Enable motion,<br/>no fault
MANUAL --> TEST: test starts
TEST --> MANUAL: test ends
MANUAL --> RESTRICTED: a restriction appears
TEST --> RESTRICTED: a restriction appears
RESTRICTED --> MANUAL: it clears
MANUAL --> DISABLED: Disable motion / STOP / fault
TEST --> DISABLED: Disable motion / STOP / fault
RESTRICTED --> DISABLED: Disable motion / STOP / fault
| State | Motion | Speed |
|---|---|---|
| DISABLED | Off. The drive is disabled and queued moves are discarded. | — |
| MANUAL | On, driven by jogs from the app | Up to Velocity Max |
| TEST | On, driven by the test program on the SD card | Up to Velocity Max |
| RESTRICTED | On, so the carriage can be backed off | Capped at Restricted Velocity |
The controller decides the state afresh on every pass, in this order:
- Any fault → DISABLED.
- Motion not requested → DISABLED.
- Any restriction → RESTRICTED.
- A test running → TEST, otherwise MANUAL.
The app doesn't show the state by name. It shows the four inputs to it, as the badges on the Control screen:

| Badges | State |
|---|---|
Fault: anything but NONE |
DISABLED |
Motion: disabled |
DISABLED |
Motion: enabled, Restriction: anything but NONE |
RESTRICTED |
Motion: enabled, Test: idle, both NONE |
MANUAL |
Motion: enabled, Test: running, both NONE |
TEST |
Faults and restrictions¶
A fault means the machine can't be trusted to move: a safety input is open, or part of the controller or its sensors has stopped. It forces DISABLED.
A restriction means the machine is somewhere it shouldn't stay: on an end stop, door open, over a force or length limit. It slows the machine and leaves it movable, because refusing to move would strand a sample under load.
Both are re-evaluated continuously from the live inputs, so they clear when their cause does. One fault is different: a servo stall latches until you disable motion, because disabling motion is what makes a stall stop being detected.
One consequence is worth knowing. Enable motion is a standing request. A fault overrides it but doesn't cancel it, so when a fault clears by itself, motion returns. Clicking Disable motion while faulted cancels the request.
Faults and Restrictions list every one.
How a test runs¶
sequenceDiagram
participant A as MaD Control
participant C as Controller
participant S as SD card
A->>C: Sample profile (limits)
A->>C: Motion program
C->>S: store program
A->>C: Enable motion, start test
loop until the program ends
S->>C: next move
C->>C: run it, measure, check limits
C->>S: record samples
C-->>A: live samples and state (when asked)
end
C-->>A: test no longer running
A->>C: Download recording
- The app turns the motion profile into a program of moves (G-code) and uploads it to the SD card, along with the sample profile's limits.
- The controller plays the program from the card, one move after another, and records force and position to the card as it goes.
- It ends the test when the program ends, when the sample passes Max force or Max displacement, or when motion is disabled for any reason.
- The app notices the test is no longer running and marks the run completed. Download data copies the recording off the card.
The app's part ends at step 1. Everything after is the controller working from its own SD card, which is why closing the tab or pulling the USB cable does not stop a test.
What the app does¶
The app only ever asks; the controller answers.
- Live data is polled about a hundred times a second, and the machine's state about once a second. Responding means answers are arriving.
- Commands (enable, jog, home, zero, start a test, write configuration) are requests the controller can refuse. A refused jog shows as a toast.
- On connect, the app asks which protocol version the firmware speaks, and sends no motion command until it matches its own. This is the firmware banner.
Two serial ports keep these roles apart. The control port carries all of the above and never resets the controller. The programming port resets the controller into its boot loader and is used only to flash firmware.
Power-up sequence¶
Power is applied in stages so the electronics and the safety chain come up in a known order:
- Turn on the mains inlet switch. This powers the controller.
- Press the momentary ON button. This applies DC power to the status switches and the emergency stop chain.
- Release the emergency stop. This applies AC power to the motor drive.
- When no fault is showing, Enable motion from the app.
To power down, reverse the order.
The controller reads its configuration at power-up, so a changed machine configuration takes effect after a restart: the app's Reboot now, or a power cycle.
Safety model¶
The machine is the safety authority
The app is never a safety device. The machine has:
- a hardware emergency stop that removes power from the drive;
- faults that switch motion off whenever a safety input opens or part of the controller stops;
- restrictions that slow it at the end stops, with the door open, or over a force or length limit;
- sample limits that end a test.
The red STOP button in the app sends a disable-motion request. It is a convenience, and it depends on the computer, the browser and the cable all working. Use the hardware emergency stop when it matters.